This laser power meter is a must for a wide array of applications, from bio-stimulation to telecommunications. These handheld units have an extremely wide flexibility and are very user-friendly.
Three detector heads are available for laser power measurement – UV (200 to 1100nm), Visible (350 to 1100nm), and Near-IR (800 to 1800nm). In addition to the photodiode heads for individual wavelengths, we offer heads for measuring the output of broadband light sources. The radiometer detector head has spectrally flat response from 400 to 1000nm and therefore can give the true total power of any broadband light source in that spectral region. The photometer head has a spectral response similar to that of the human eye and can therefore make measurements in eye response units of Lux. The photometer head is designed with a small detector where the source overfills the detector. It measures the light intensity per unit area in units of Lux or Foot Candles.
The laser power meter recognizes the pre-calibrated detector head when it is attached. Detector heads can be oriented over a 180° angle and come with an attachable 0.5" dia. x 2.25" L post, 0.875" OD / 0.5" ID post holder, and 3" x 4" sliding base for breadboard mounting. Note: Socket Head Cap Screws not included. Heads also include removable calibration filter.
Functions of the power meter are controlled via menu options on the LCD display that can be controlled by the “Menu” button or three “soft” buttons below the LCD. User navigates through desired menus to select various pre-defined operations. The High Performance version (#58-359) offers the ability to store up to 54,000 data points on-board in 10 files, a high-resolution screen with a selectable digital or analog display, and several additional features.
Meter housing is molded high-impact plastic with kickstand and EMI conductive shielding to allow use in proximity to pulsed lasers. On/off switch on left side. Supertwist LCD display has switchable electroluminescent backlight (active from power charger only), contrast adjustment, and large 12mm digits. User manual and battery charger included. Unit also comes with N.I.S.T. traceable certificate of calibration in accordance with ISO 10012-1.
Optional RS232 adapter for standard meter #55-642 allows a 9600-baud connection between the laser power meter and the user’s computer via a 1.5m long cable. Interface software is included and is LabVIEW® compatible.
Note: Complete system requires a detector head and laser power meter #55-642 or #58-359.
| Laser Power Meter Specifications | ||
| Standard Model | High Performance Model | |
| Wavelengths | Selectable (within 1nm) | Selectable (within 1nm) |
| Refresh Time | 15 times/sec | 15 times/sec |
| Analog Output | 1 volt full scale | 1, 2, 5, & 10 volt full scale |
| Overall Dimensions | 203mm L x 95mm W x 38mm D | 208mm L x 117mm W x 40mm D |
| LCD Display Dimensions | 122 x 32 pixels, 12mm digits | 320 x 240 pixels, 18mm digits |
| Weight | 2 lbs. | 2 lbs. |
| Battery | Rechargeable NiCad 12V DC charger (included) also acts as wall adapter | Rechargeable NiCad 12V DC charger (included) also acts as wall adapter |
| Lifetime | 18 hours of continuous use between charges | 18 hours of continuous use between charges |
| UV/VIS/NIR Heads - Common Specs | |
| Beam Position Dependence | ±2% across detector area |
| Response Time | 0.2 sec |
| Damage Threshold | 10W/cm2 (without filter), 50W/cm2 (with filter) |
| Weight | 550g |
| Dimensions | |
| Handle | 68.5mm L x 10mm diameter |
| Head | 40.5mm L x 18mm W x 10.2mm D (without filter) 42mm L x 21.5mm W x 12.8mm D (with filter) |
| UV/VIS/NIR Detector Heads | |||
| Head Type | UV Detector Head | Visible Detector Head | Near-IR Detector Head |
| Wavelength Range | |||
| Without Filter | 200-1100nm | 350-1100nm | 800-1800nm |
| With Filter | 220-1100nm | 400-1100nm | 800-1800nm |
| Power Scales | |||
| Without Filter | 3.000mW, 300.0µW, 30.00µW, 3.000µW, 300.0nW, 30.00nW, 3.000nW or dBm | 30.00mW, 3.000mW, 300.0µW, 30.00µW, 3.000µW, 300.0nW, 30.00nW or dBm | 30.00mW, 3.000mW, 300.0µW, 30.00µW, 3.000µW, 300.0nW or dBm |
| With Filter | 300.0mW, 30.00mW, 3.000mW, 300.0µW or dBm | 300.0mW, 30.00mW or dBm | 300.0mW, 30.00mW or dBm |
| Max Pulse Energy | |||
| Without Filter | 0.4µJ | 2µJ | 2µJ |
| With Filter | 15µJ | 20µJ | 20µJ |
| Aperture | 10mm x 10mm | 10mm x 10mm | 5mm Diameter |
| Noise | ±1pW at mid spectrum without filter | 0.02nW at mid spectrum without filter | 0.2nW at mid spectrum without filter |
| Accuracy | |||
| Without Filter | 200 to 250nm: ±5.0% 250 to 950nm: ±3.0% 950 to 1100nm: ±5.0% |
350 to 400nm: ±10.0% 400 to 950nm: ±5.0% 950 to 1100nm: ±7.0% |
800 to 1600nm: ±5.0% 1600 to 1800nm: ±7.0% |
| With Filter | 220 to 300nm: ±10.0% 300 to 950nm: ±5.0% 950 to 1100nm: ±7.0% |
400 to 950nm: ±5.0% 950 to 1100nm: ±7.0% |
800 to 1600nm: ±7.0% 1600 to 1800nm: ±7.0% |
| Background Subtraction | N/A | 95-98% of background is cancelled automatically under normal room conditions | N/A |
| Maximum Power vs. Wavelength for UV/Visible/Near-IR Detector Heads | ||||||||
|
UV Detector Head
|
Visible Detector Head
|
Near-IR Detector Head
|
||||||
| Wavelength | without Filter | with Filter | Wavelength | with out Filter | with out Filter | Wavelength | with out Filter | with Filter |
| 250-350nm | 3mW | 300mW | <488nm | 30mW | 300mW | 800nm | 12mW | 120mW |
| 400nm | 3mW | 300mW | 633nm | 20mW | 300mW | 1000-1300nm | 30mW | 300mW |
| 600nm | 3mW | 300mW | 670nm | 13mW | 200mW | 1400nm | 30mW | 250mW |
| 800-950nm | 2.5mW | 150mW | 790nm | 10mW | 100mW | 1500nm | 25mW | 80mW |
| 1064nm | 3mW | 300mW | 904nm | 10mW | 150mW | 1600nm | 30mW | 100mW |
| — | — | — | 1064nm | 25mW | 250mW | 1800nm | 30mW | 300mW |
| Radiometer and Photometer Heads | ||
|
Head Type
|
Radiometer Head
|
Photometer Head
|
| Wavelength Range | 400 - 1000nm | 400 - 700nm |
| Power/Measurement Scales | 8.000mW, 800.0µW, 80.00µW, 8.000µW, 800.0nW, 80.00nW, 8.000nW or dBm | 200.0K Lux, 20.00K Lux, 2.000K Lux, 200.0 Lux, 20.00 Lux, 2.000 Lux or Footcandles |
| Damage Threshold | 10W/cm2 | 10W/cm2 |
| Max Power vs. Wavelength | 6mW (all wavelengths) | — |
| Maximum Pulse Energy | 1µJ | — |
| Aperture | 10mm x 10mm | 2.4mm x 2.8mm |
| Response Time | 0.2 sec | 0.2 sec |
| Beam Position Dependence | ±2% | — |
| Noise Level | ±2pW at mid spectrum | ±1m Lux at mid spectrum |
| Background Subtraction | N/A | N/A |
| Accuracy | Worst case deviation from flat spectrum ±10% (See Graph) | See Graph |
| Weight | 550g | 550g |